Forum Post: RE: ADS1296: Continuous data mode sometimes only returns 0s

$
0
0
Yes. The input signal is always the same. Digging further, I actually found that my issue is exactly the same as Mich in this post: e2e.ti.com/.../470033. It looks like the device doesn't take into account the RDATAC cmd send before conversions start. If I send a single CMD_RDATA before each data read, then I don't have any issues...