Quantcast
Viewing all articles
Browse latest Browse all 27703

Forum Post: RE: ADS1296: Continuous data mode sometimes only returns 0s

Yes. The input signal is always the same. Digging further, I actually found that my issue is exactly the same as Mich in this post: e2e.ti.com/.../470033. It looks like the device doesn't take into account the RDATAC cmd send before conversions start. If I send a single CMD_RDATA before each data read, then I don't have any issues...

Viewing all articles
Browse latest Browse all 27703

Trending Articles



<script src="https://jsc.adskeeper.com/r/s/rssing.com.1596347.js" async> </script>