Forum Post: RE: ADS1294R: Bits Flickering in Lead Off registers

$
0
0
Hi Smith, Initially, the machines are working well for 2 to 3 months after production, after that in 5 out of 50 machines we observed like that. The Short observed in ADS1294R . The Board is working fine when we replaced with new ADS1284R IC. What could be the reason for this problem? Did you verify the Register Configurations which i shared in the last post? Hope the configurations are fine. Kindly confirm the same to verify the Lead off Detection Bits flickering issue. Thanks and Regards, Vignesh N