Forum Post: RE: ADS1293: ADS1293 3-Lead ECG noise

$
0
0
Hi Alexander As below waveform is using patient simulator & output sine wave for you refer. 1. lead-off (disable) 2.lead-off (8nA) About your recommend RC filter for electrode inputs . May I refer electrode input RC filter of ADS1298EVM? or do you have other recommend RC filter circuit? As below waveform is using ADS1298EVM RC filter to measure human body. not use patient simulator. 1.lead-off (disable) 2.lead-off (8nA) Do you have any suggest to work on to further reduce the noise?