Forum Post: RE: Questions about TX-SDK-V2 EVB

Hi Friday,Sorry you can't see the schematic. Let's try again. The schematic is not on the LM96511 datasheet. It's part of our internal demo board. As you can see from the schematic, the external diode...

View Article


Forum Post: RE: Questions about TX-SDK-V2 EVB

Hi Charles, any news for the power supply release ?Best regards.

View Article


Forum Post: RE: AFE5808

Chandu,The purpose of the VCNTL pins are for Time-Gain Compensation (TGC), where the gain of the VCAT block is varied over time.  This can be used in the ultrasound system to offset natural attenuation...

View Article

Forum Post: RE: Questions about TX-SDK-V2 EVB

Hi Olivier,We are hoping we can release the power module schematic to web later this week.Regards,Charles

View Article

Image may be NSFW.
Clik here to view.

Forum Post: RE: Ultrasound transducer interface to pulser outputs

Hi,  Actually I have designed individually  customized boards for the pulser(lm96550), Beamformer (lm96570), tx/rx switch (tx810) with reference of Tx_sdk_v1.    The below Image shows the beamformer...

View Article


Forum Post: RE: AFE5808

For medical diagnosis to get Time gain compensation, Can you suggest me VCNTL analog signal. I think it is ramp, but I can't understand what should be the frequency. On what criteria the analog signal...

View Article

Forum Post: AFE5809 Speed Grade

Hi,Could someone please provide for me the speed grade of the AFE5809?I need this to set up the ALTPLL on my Cyclone V.Also are there any examples on LVDS with the AFE5809?Thanks!Jack

View Article

Forum Post: RE: Ultrasound transducer interface to pulser outputs

Hi Chandu,Does the pulser output get less distorted when you set B3B2B1 to 000 or 001? I am not familiar with the TX810. However, looking at fig 14 of the TX810 datasheet, it might loading the output...

View Article


Forum Post: RE: multi AFE5808

Hi Ly,1. Why not? Sure you can2. Vcntl input resistance of AFE5808 is 200kOhm. Connecting 32 devices in parallel gives you 6250 Ohm.    Max TGC voltage is 1.5V, therefore you require 240uA. Max output...

View Article


Forum Post: RE: AFE5809 Speed Grade

Hi Jack,AFE5809 is 65MSPS. Regarding the LVDS I understand you mean deserializer. I'm not an expert on Altera FPGA's, I use Xilinx. I think you should ask this question on Altera forum. However, take a...

View Article

Forum Post: RE: Ultrasound transducer interface to pulser outputs

Hi Chandu,You don't need any special cable interfacing to the output of the pulser. Is it possible you can post the schematic of your design?Regards,Charles

View Article

Forum Post: RE: Ultrasound transducer interface to pulser outputs

Hi Chandu,1) Remove  220pF capacitor.2) Remove 2.2K resistor.3) Connect the pulser output to the transducer using a 50Ohm well shielded cable.4) Now observe the signal at input and output of the T/R...

View Article

Forum Post: RE: multi AFE5808

Hi Mariusz,Thanks for your help and time.Best Regards,Ly Nguyen

View Article


Forum Post: RE: AFE5808

Mariusz, thanks for answering the quesitons! i agree with you that TGC frequency is not a common term. Our AFE5808/9 datasheets use pulse repetition frequency (PRF) since it is a more common term in...

View Article

Forum Post: Connecting inputs of TX810

Hi,Currently we're designing a small ultrasound device for development and research. We are interested in using TX810. In our system high-channel count or power consumption aren't most important...

View Article


Forum Post: Need help on project based on TI's portable ultrasound

HiIam doing my final year project based on the TIs portable ultrasound at http://www.ti.com/solution/ultrasound_system_portable. But i have some questions :- Since the TI's system is of 8 channel what...

View Article

Forum Post: RE: need help on 16 channel US pulser

Hi Naftali,Are you using the LM96551 or the LM96550 pulser? If you are using the LM96551, the +/-40V is built it already. If you are using the LM96550, please use the power up sequence on page 10 of...

View Article


Forum Post: RE: Connecting inputs of TX810

Damian,I believe it is possible to connect several channels in parallel.  Is that what you are trying to do?  Inputs shorted together and outputs shorted, as well?  You can reduce Ron by doing so, but...

View Article

Forum Post: RE: Ultrasound transducer interface to pulser outputs

Hi Mariusz,If you look at Chandu's post on Dec. 18, he mentioned just when T/R switch is NOT connected to the output of the pulser, he saw no visible distortion. However, when he connect the T/R switch...

View Article

Forum Post: RE: Ultrasound transducer interface to pulser outputs

Hi,   previously when I am using switch and pulser boards seperatly, pulser o/p are connected to switch i/p and transducer using normal female connectors then I got the distortion problem.   I planned...